IET/QUADTECH 1910/1920 1 MHZ LCR METER

Dich vụ sửa chữa, bảo dưỡng, cải tiến các hệ thống dây chuyền tự động công nghiệp

Data Sheet

Why Buy The 1920 Precision 1 MHz LCR Meter?

  • Robust and production oriented LCR meter for automating your impedance measurements
  • A wide frequency range 20 Hz to 1 MHz at an affordable price
  • When both ac impedance and dc resistance measurements are needed
  • Standard RS232, IEEE-488 & PLC interfaces simplify automation
  • When internal bias current is required for testing coils and inductors
  • Configure instrument and log results to PC using the 1900 Virtual Front Panel or LabView driver

Description

 

There are two versions of the 1 MHz Precision LCR Meter, the 1910 and 1920 


The 1920 1 MHz LCR Meter is the most popular and designed to perform capacitance and impedance measurements on a variety of electronic components and materials over a frequency range from 20 Hz to 1 MHz. The 1920 features programmable bias voltage from 1 mV to 2 V for biasing capacitors and other semiconductor devices during measurements.

The 1910 Inductance Analyzer has the same feature as the 1920, with the bias voltage feature being replace by a programmable bias current of up to 1 A. The bias current features makes the 1910 designed to perform inductance and impedance measurements on coils and wire wound devices over a frequency range from 20 Hz to 1 MHz.

Emphasis has been placed on ease of use through menu programming, and fast accurate production testing via its remote control capability with RS-232, handler and GPIB interfaces.

15 Different Impedance Parameters/20 Total Parameters Measure and display any two parameters simultaneously, to achieve coverage and flexibility not previously available.

Automatic Test Sequencing Run up to six different tests in sequence with a single push of the start button. Each test can have different conditions and limits.
Program Storage Test setups can be stored and recalled from internal memory. The front panel can be locked out, with password protection, to ensure procedures are run the same way every time.
Configure instrument and log results to PC IET offers a Window based program that can be downloaded for free. 
1900 Virtual Front Panel

If you are looking for a more cost effective or higher preforming LCR Meter such as the IET / QuadTech 7600 Precision LCR Meter or IET / GenRad 1689 Digibridge, See our LCR Meter Selection Guide

1910/1920 Basic Specifications

(See datasheet for full specifications)

Measured Parameters and Basic Measurement Accuracy:

Parameter Measurement Range Fast Medium Slow
Ls,Lp 0.001 nH to 99.999 H +/-0.5% +/-0.25% +/-0.1%
Cs,Cp 1 pF to 9.9999 F +/-0.5% +/-0.25% +/-0.1%
DF 0.00001 to 99.999 +/-0.005 +/-0.0025 +/-0.001
Q 0.00000 to 9999.9 +/-0.005 +/-0.0025 +/-0.001
Y,Gp,Bp 10 nS to 9999.9 S +/-0.005 +/-0.0025 +/-0.001
|Z|,Rs,Rp,Xs,ESR 0.0001 mΩ to 99.999 MΩ +/-0.5% +/-0.25% +/-0.1%
Phase Angle -180.00 to +179.99 degrees +/-1.8° +/-0.9° +/-0.18°
DCR 0.1 mVto 100 KV +/-0.5% +/-0.25% +/-0.1%
DUT AC Voltage 20 mV to 1.0 V +/-2% +/-2% +/-2%
DUT AC Current 1 μA to 150 mA +/-2% +/-2% +/-2%
DUT DC Voltage 20 mV to 1.0 V +/-2% +/-2% +/-2%
DUT DC Current 1 μA to 150 mA +/-2% +/-2% +/-2%

*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty. Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q) Factors, Impedance Z , Admittance Y , Phase Angle (0-), Equivalent Series Resistance (ESR), Conductance (Gp), Reactance (Xs), Susceptance (Bp) Note: s = series, p = parallel, ESR equivalent to Rs
Test Frequency Range: 20 Hz to 1 MHz, Continuous
Test Frequency Resolution: 1 Hz from 20 Hz to 1 kHz, 4 digits >1 kHz
Test Frequency Accuracy: +/- (0.01% +0.02 Hz)
Measurement Speed: Fast: 40 meas/sec Medium: 8 meas/sec Slow 1 meas/sec
Ranging: Automatic,Range Hold or user selectable
Trigger: internal (automatic), External(via RS-232,IEEE-488.2, Handler interfaces or Manual
Source Impedance: 5, 25, 50, 100 ohms selectable
AC Test Signal Voltage: 20 mV to 1.0 V (open circuit) in 5 mV steps
DC Bias Current Internal: 0.0 to 1.0 A in 1 mA steps External: up to 20.0 A
Standard Interfaces: IEEE-488.2 • RS-232 • Handler • Serial Printer
Measurement Delay: Programmable from 0 to 1000 ms in 1 ms steps
Averaging: Programmable from 1 to 1000
Median Value: Averaged over last three measurements
Built-in automatic adjustment procedure
Test Terminals: Front panel, four terminal (BNC) Optional Test Fixtures Available
Dimensions: 17x 5.25x 16in (432x 133x 406 mm)
Weight: 15lbs (8kg) net, 21lbs (9.9kg) shipped