hprs-series-high-power-resistance-decade-box.png

HPRS SERIES HIGH POWER RESISTANCE DECADE BOX

Dich vụ sửa chữa, bảo dưỡng, cải tiến các hệ thống dây chuyền tự động công nghiệp

Data Sheet

The High-Power Decade Resistance Substituter (HPRS) Series is a family of resistance decades offering a broad choice of high-power, excellent-performance resistance sources.

The HPRS Series resistance decades employ state-of-the-art precision resistors of various types for high accuracy, high stability, and low temperature and power coefficients of- resistance. High-power resistors within the unit are made available without sacrificing other electrical properties.

The standard models offer a choice of 3 to 8 decades. The panels are clearly labeled showing the step size and maximum voltage and current limitations for each decade.

With a resolution as low as 1 mΩ and a maximum resistance of 10 MΩ , the HPRS resistance decade box may be used as a general purpose substituter as well as a high power load for testing power supplies and for other high-power applications.

Applications include calibration of meters and instruments. HPRS resistance decades are useful development tools wherever precise resistances with high-power handling capacity are required.

HPRS Series High Power Resistance Decade Basic Specifications

(See datasheet for full specifications) 

Accuracy:
Option F: ±(1% + 20 mΩ)
Option C: ±(0.5% + 20 mΩ) 

Zero resistance: <5 mΩ per decade

Maximum voltage to case: 1000 V peak 

Switches: continuous rotation 10 positions marked "0"-"9"

Resistance per step Total decade resistance Max current* Max power per step* (W) Temperature coefficient (ppm/°C) Resistor Type
1 mΩ 0.009 Ω 8 A 0.036 ±50 Resistance Wire
10 mΩ 0.09 Ω 6 A 0.36 ±50
100 mΩ 0.9 Ω 6 A 3.6 ±20
1 Ω 9 Ω 5 A 25 -20 to +80 Power Film
10 Ω 90 Ω 1.5 A 25 -20 to +80
100 Ω 900 Ω 0.5 A 25 -20 to +80
1 kΩ 9 kΩ 150 mA† 25† -20 to +80
10 kΩ 90 kΩ 50 mA† 25† -20 to +80
100 kΩ 900 kΩ V limit† V limit† -20 to +80 Wirewound/metal film
1 MΩ 9 MΩ V limit† V limit† ±50 Wirewound/film
           
*Subject to maximum of 250 W max. per unit: 9 Steps per decade, multiply the decade value by the power per step. Multiple decades can be used to share power loads
†Subject to 1000 V (dc+ac) peak max